當(dāng)前位置:主頁>產(chǎn)品展示>英文資料
BJ-Ⅱ(譯)
發(fā)布時間:2017-02-16BJ-Ⅱ Disintegration Time Limit Tester
Disintegration time limit tester is a drug test instrument to test the disintegration time limit of tablets, capsules and pills in accordance with Chinese Pharmacopoeia.
Technical indexes:
△ Temperature prefabrication range: room temperature:~50℃, display resolution:0.1℃.
△ Temperature control accuracy: ±0.3℃
△ Timing prefabrication range: 1min~900min
△ Time control accuracy: ±0.5min
△ Work noise: <60db
△ Number of lifting baskets: 2 sets
△ Lifting frequency of basket: (30~31) times/min.
△ Lifting distance of basket: (55±1) mm
△ Minimum spacing from screen to beaker bottom: 25mm±2mm
△ Mesh size: standard configuration 2mm, (optional for 0.425mm and 1mm bore diameters)
△ Power supply:220V50Hz
△ Total power:600W
△ Overall dimensions:L*W*H 400mm*320mm*440mm
Main features:
◎ Double-basket working position system measures disintegration time limit.
◎ 2 sets of baskets, which can conduct independent disintegration experiment respectively.
◎ Electronic temperature sensor can display and monitor various points in water bath case and the temperature in the beaker. High-precision digital electronic sensor is adopted to ensure the high precision and high accuracy of water bath temperature without the need for calibration.
◎ The instrument controls the bath temperature at 37.0℃ automatically (prescribed by Pharmacopeia). It can also reset the preset temperature again at any time.
◎ The instrument sets the lifting time of baskets to be 15min automatically, which can also be reset at any time.
◎ It adopts the computer control technology which is centered on SCM to control two performance parameters including bath temperature and working time intelligently.
◎ The instrument can monitor bath temperature, give over-temperature alarm and protect itself automatically.